12 A-96.250.871 / 120417
AMI CACE
Product Description
2.2. Instrument Specification
Power Supply Voltage: 100240 VAC (± 10%)
50/60 Hz (± 5%)
or 24 VDC (± 10%)
Power consumption: max. 30 VA
Sample
requirements
Flow rate: 34 l/h
Temperature: up to 50 °C
Inlet pressure: up to 0.5 bar
Outlet pressure: pressure free
Use of a SWAN Backpressure Regulator is highly recommended.
Particle filtration recommended in case of high iron concentration.
NOTICE: No oil, no grease, no sand. Use of film forming
products may reduce lifetime of EDI module.
On-site
requirements
The analyzer site must permit connections to:
Sample inlet: Swagelok 1/4” adapter for stainless
steel tube
Sample outlet: G 3/8” adapter for flexible tube
diam. 20 x 15 mm
Measuring
range
Measuring range Resolution
0.055 to 0.999 S/cm 0.001 S/cm
1.00 to 9.99 S/cm 0.01 S/cm
10.0 to 99.9 S/cm 0.1 S/cm
100 to 1000 S/cm 1 S/cm
Automatic range switching.
EDI capacity SC
max
= 40 S/cm as NH
4
OH
SC
max
= 350S/cm as NaOH
Accuracy ±1% of measuring value or ±1 digit (whichever is greater)
Electronics
housing
Aluminum with a protection degree of IP 66 / NEMA 4X
Ambient temperature: -10 to +50 °C
Humidity: 1090% rel., non condensing
Display: backlit LCD, 75 x 45 mm
Dimensions Panel: Dimensions: 280 x 850x 200 mm
Screws: 8 mm diameter
Weight: 14 kg