AMI CACE
Program List and Explanations
A-96.250.871 / 120417 69
The relay will be activated repetitively depending on the pro-
grammed time scheme.
5.3.2.24 Mode: Operating mode (interval, daily, weekly)
5.3.2.340 Interval/Start time/Calendar: Dependent on options operating
mode.
5.3.2.44 Run time: time the relay stays active.
Range: 5–32’400 Sec
5.3.2.54 Delay: during run time plus the delay time the signal and control
outputs are held in the operating mode programmed below.
Range: 0–6’000 Sec
5.3.2.6 Signal Outputs: select the behavior of the signal outputs when the
relay closes. Available values: cont., hold, off
5.3.2.7 Output/Control: select the behavior of the control outputs when the
relay closes. Available values: cont., hold, off
The relay will be switched via the Profibus input. No further param-
eters are needed.
5.3.2.1 Function = Timer:
5.3.2.1 Function = Fieldbus: