AMI CACE
Maintenance
A-96.250.871 / 120417 37
6.5. Longer Stop of Operation
If the instrument is not used for a longer period of time (2 months or
more), drain the EDI module and seal the tube fittings marked in
red using the end caps [A] supplied with the instrument.
Procedure 1 Stop sample flow.
2 Unscrew the top ends of tubes 1 and 2.
3 Drain the EDI module through tube 2.
4 Seal tubes 1 and 2 using the end caps [A].
5 Unscrew tubes 3, 5 and 10 at the positions marked in red and
seal them using the end caps [A].
6 Shut off power of the instrument.
A End cap