AMI CACE
Program List and Explanations
A-96.250.871 / 120417 51
2.2.2 Miscellaneous:
2.2.2.1 Case Temp: Shows the current temperature in [°C] inside the trans-
mitter.
2.2.3 EDI:
2.2.3.1 Actual current: Current in mA applied to the EDI module.
Actual voltage: Resulting voltage in mV.
Total cu r rent: Amount of electric charge in Ah since the last ex-
change of the EDI module.
Total fl o w: Amount of sample water in L since the last exchange of
the EDI module.
Last exchange: Date of the last exchange.
2.3 Sample
2.3.1 Sample ID: Shows the identification assigned to a sample. This
identification is defined by the user to identify the location of the
sample.
2.3.2 Sample Flow: Shows the current sample flow in l/h and the Raw
Value in Hz.
The Sample flow must be above 2.5 l/h.
2.3.3 Sample Temp:
2.3.3.1 Temp 1: Shows the current sample temperature at sensor 1 in °C.
(Pt 1000): Shows the current temperature at sensor 1 in Ohm.
Temp 2: Shows the current temperature at sensor 2 in °C.
(Pt 1000): Shows the current temperature at sensor 2 in Ohm.
2.4 I/O State
Shows current status of all in- and outputs.
2.4.1/2.4.2
2.5 Interface
Only available if optional interface is installed.
Review programmed communication settings.
Alarm Relay: Active or inactive.
Relay 1/2: Active or inactive.
Input: Open or closed.
Signal Output 1/2: Actual current in mA
Signal Output 3:(optional) Actual current in mA