40 A-96.250.871 / 120417
AMI CACE
Troubleshooting
Error Description Corrective action
E001
Cond. 1 Alarm high – check process
– check programmed value, see 5.3.1.1,
p. 62
E002
Cond. 1 Alarm low – check process
– check programmed value, see 5.3.1.1,
p. 62
E003
Cond. 2 Alarm high – check process
– check programmed value, see
5.3.1.1.2.1, p. 63
E004
Cond. 2 Alarm low – check process
– check programmed value, see
5.3.1.1.2.25, p. 63
E007
Temp. 1 Alarm high – check process
– check programmed value, see
5.3.1.1.4, p. 63
E008
Temp. 1 Alarm low – check process
– check programmed value, see
5.3.1.1.4, p. 63
E009
Sample Flow high – check sample inlet pressure
E010
Sample Flow low – check sample inlet pressure
– check if the following components are
clogged:
inlet filter (if installed)
tubes
EDI module
– If necessary, replace clogged parts.
See Tube numbering, p. 43 and
Replacing the EDI module, p. 44.
E011
Temp. 1 shorted – Check wiring of temperature sensor
– Check temperature sensor
E012
Temp. 1 disconnected – Check wiring of temperature sensor
– Check temperature sensor
E013
Case Temp. high – check case/environment temperature
– check programmed value, see
5.3.1.4.1, p. 65