56 A-96.250.871 / 120417
AMI CACE
Program List and Explanations
5.1.2.2.5 Temp. comp:
5.1.2.2.5.1 Comp.: Available compensation models:
Strong acids
5.2 Signal Outputs
NOTICE: The navigation in the menu <Signal Output 1> and
<Signal Output 2> is equal. For reason of simplicity only the
menu numbers of Signal Output 1 are used in the following.
5.2.1 Signal Output 1: Assign process value, the current loop range and
a function to each signal output.
5.2.1.1 Parameter: Assign one of the process values to the signal output.
Available values:
Cond 1 (cc)
Cond 2 (sc)
Tem p . 1
Tem p . 2
Difference
Sample flow
pH
Ammonia
5.2.1.2 Current Loop: Select the current range of the signal output. Make
sure the connected device works with the same current range.
Available ranges: 0–20 mA or 4–20 mA
5.2.1.3 Function: Define if the signal output is used to transmit a process
value or to drive a control unit. Available functions are:
Linear, bilinear or logarithmic for process values.
See As process values, p. 57
Control upwards or control downwards for controllers.
See As control output, p. 59