Program List and Explanations
5.3.1.1.5 Ammonia (if Calculations = yes)
5.3.1.1.5.1 Alarm High: If the measured value rises above the alarm high val-
ue, the alarm relay is activated and E035 is displayed in the mes-
sage list.
Range: 0.00–500 ppm
5.3.1.1.5.25 Alarm Low: If the measured value falls below the alarm low value,
the alarm relay is activated and E036 is displayed in the message
list.
Range: 0.00–500 ppm
5.3.1.1.5.35 Hysteresis: Within the hyst. range, the relay does not switch. This
prevents damage of relays contacts when the measured value fluc-
tuates around the alarm value.
Range: 0.00–500 ppm
5.3.1.1.5.45 Delay: Duration, the activation of the alarm relay is retarded after
the measuring value has risen above/fallen below the programmed
alarm.
Range: 0–28‘800 Sec
5.3.1.2 Sample Temp.
5.3.1.2.1 Temp. 1
5.3.1.2.1.1 Alarm High: If the measured value rises above the alarm high val-
ue, the alarm relay is activated and E007, is displayed in the mes-
sage list.
Range: 30–200 °C
5.3.1.2.1.25 Alarm Low: If the measured value falls below the alarm low value,
the alarm relay is activated and E008 is displayed in the message
list.
Range: -10 to +20 °C
5.3.1.2.2 Temp. 2
5.3.1.2.2.1 Alarm High: If the measured value rises above the alarm high val-
ue, the alarm relay is activated and E037, is displayed in the mes-
sage list.
Range: 30–200 °C
5.3.1.2.2.25 Alarm Low: If the measured value falls below the alarm low value,
the alarm relay is activated and E038 is displayed in the message
list.
Range: -10 to +20 °C