5.4 Miscellaneous
5.4.1 Language: Set the desired language.
Available settings: German /English/French/Spanish/Italian
5.4.2 Set defaults: Reset the instrument to factory default values in three
different ways:
Calibration: Sets calibration values back to default. All other
values are kept in memory.
In parts: Communication parameters are kept in memory. All
other values are set back to default values.
Completely: Sets back all values including communication
parameters.
5.4.3 Load Firmware: Firmware updates should be done by instructed
service personnel only.
5.4.4 Password: Select a password different from 0000 to prevent unau-
thorized access to the menus “Messages”, “Maintenance”, “Opera-
tion” and “Installation”.
Each menu may be protected by a different password.
If you forgot the passwords, contact the closest SWAN representa-
tive.
5.4.5 Sample ID: Identify the process value with any meaningful text,
such as KKS number.
5.4.6 Line Break Detection: Define if message E028 should be issued in
case of a line break on signal output 1 or 2.
Choose between <Yes> or <No>.