EasyManua.ls Logo

SYSMEX CS-5100 LAS - 2.3 Disconnecting the LAS; 2.4 Shutting down

SYSMEX CS-5100 LAS
32 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 2 Basic Operations
CS-5100 LAS Instructions for Use
Revised October 2014
2-2
2.3 Disconnecting the LAS
Follow the procedure below to disconnect the CS-5100 and LAS before analyzing a sample with CS-5100 by itself
without any LAS connection, replacing the reagent, or discarding the waste fluid from the instrument.
1 On the IPU status bar, press the LAS setting button of the LAS connection
status display.
( P.2- 1 “Chapter 2: 2.2 Checking the LAS connection status”)
The dialog box on the right appears.
2
Tick the [Suspend Processing of LAS Samples and use the instrument] check
box and press [OK].
The CS-5100 and the LAS are now disconnected and the dialog box closes.
* To reconnect the CS-5100 and the LAS, remove the [Suspend Processing of LAS Samples and use the
instrument] checkbox in the same dialog box and press [OK].
2.4 Shutting down
To shut down the instrument properly, refer to the procedures described in the CS-5100 Instructions for Use.
( CS-5100 Instructions for Use “Chapter 3: 3.4.6 Shutdown”)
LAS setting button Disconnects or reconnects the connection between the CS-5100 and the
LAS.
( P. 2 - 2 “Chapter 2: 2.3 Disconnecting the LAS”)
Instrument control
status
Lights orange or turns OFF depending on the control status of the
instrument.
Indicator Control status
Lit The analysis of the sample on the tracks is disabled.
The LAS is disconnected by the LAS setting button.
OFF The instrument is ready to analyze a sample on the
tracks.

Related product manuals