EasyManua.ls Logo

SYSMEX CS-5100 LAS - Chapter 6 Setting Up the System

SYSMEX CS-5100 LAS
32 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 6 Setting Up the System
CS-5100 LAS Instructions for Use
Revised October 2014
6-1
Chapter 6 Setting Up the System
This chapter explains the system settings of the instrument that can be set only when the CS-5100 is
connected to the LAS.
See the CS-5100 Instructions for Use for more details on the other system settings.
( CS-5100 Instructions for Use “Chapter 4: 4.1 Setting up the system”)
6.1 [Sample Volume Check] screen
When the LAS is connected to the instrument, the following items appear on the [Sample Volume Check] screen
([Operational Settings] [Sample Volume Check] in the [System Settings] dialog box).
See the CS-5100 Instructions for Use for more details on the [Sample Volume Check] screen.
( CS-5100 Instructions for Use “Chapter 4: 4.1.7.6 [Sample Volume Check] screen”)
[Sample Volume Check] screen
[LAS] When checking the sample volume with respect to the sample on the tracks, tick this
checkbox to set the range of sample volume (liquid level).

Related product manuals