Chapter 5 Performing QC Analysis
CS-5100 LAS Instructions for Use
Revised October 2014
5-1
Chapter 5 Performing QC Analysis
This chapter explains the QC analysis when operating the CS-5100 using the tracks.
5.1 Performing the QC analysis with the control set on the
CS-5100 before analyzing a sample on the tracks
Before analyzing a sample on the tracks, disconnect the CS-5100 from the LAS and perform the QC analysis with the
control that is set on the CS-5100.
1 Disconnect the CS-5100 from the LAS.
( P. 2 - 2 “Chapter 2: 2.3 Disconnecting the LAS”)
2
Set the control on the CS-5100 to perform QC analysis.
(CS-5100 Instructions for Use “Chapter 3: 3.10 Performing QC analysis”)
Caution!
Follow the instructions attached to each control and reagent when using these substances.