Chapter 4 Checking Analysis Results
CS-5100 LAS Instructions for Use
Revised October 2014
4-1
Chapter 4 Checking Analysis Results
This chapter explains how to check the analysis results of the track sample.
4.1 [Joblist] screen
The analysis results of the track sample can be reviewed on the [Joblist] screen.
See the CS-5100 Instructions for Use for more details on the [Joblist] screen.
(➤ CS-5100 Instructions for Use “Chapter 3: 3.9.2.1 [Joblist] screen”)
For the analysis results of the track sample, the [Rack No. - Pos.] field will appear on the [Joblist] screen as follows:
4.2 Displaying the analysis results of the track sample only
You can specify the display conditions for the [Joblist] screen and display the analysis results only of track sample.
1 Register the display conditions through the [Display Condition] dialog box for
the [Joblist] screen.
The following procedures explain how to set the display to only show the analysis results of the track sample.
(1) Press the [Joblist] icon on the menu screen or the [Joblist] button on the toolbar.
The [Joblist] screen appears.
(2) Press [Customize] on the operation panel.
The [Customize] dialog box appears.
* If [Customize] is not displayed, press [More] to switch the page.
(3) Press the [Display Condition] tab.
(4) Press [Add].
The [Filter / Sort / View] dialog box appears.
(5) Press the [Filter] tab.
(6) Enter the display condition name in the [Display Condition] field.
(7) Tick the [Use Filter] checkbox.
(8) Select both [Specify Sample Distinction Code.] and [LAS Sample].
Remove all other checkboxes.
(9) Press [Save].
The [Filter / Sort / View] dialog box closes and the conditions are saved.
[Rack No. - Pos.] The number is a sequential number with [L] at the beginning, which is obtained after
turning ON the IPU (e.g. L00001).