Chapter 3 Performing Analysis
CS-5100 LAS Instructions for Use
Revised October 2014
3-3
3.2.2 Analyzing the STAT sample while analyzing the sample on the
tracks
Through the host inquiry or manual order registration, you can also analyze a STAT sample while analyzing
another sample on the tracks.
See the CS-5100 Instructions for Use for more details on the STAT sample analysis.
(➤ CS-5100 Instructions for Use “Chapter 3: 3.8.7 Analyzing STAT samples”)
1 Set a STAT sample to the CS-5100.
2
Register the order.
An order registration is not required if you are analyzing through the host inquiry.
3
Close the STAT/buffer table cover and press the start button.
Or press the [Start] button on the IPU toolbar.
This halts the analysis on the tracks and proceeds the analysis on the STAT sample.
After aspiration of the STAT sample is completed, the analysis on the tracks resumes automatically.