V/D
S
ERIES
P
ARAMETERS
8-2 Revision: 2.1
8.1 Access level
Do not give any codes for access levels to unauthorised
persons!
In particular, the code for access level 4 is exclusively
reserved for service technicians and must never be passed
on to the operator.
Parameter settings are made after entering the code for the necessary
access level. The following access levels are available.
Access level Code Target group
1 1000 Users
2 0011 Option for the operator to offer
different levels of access (e.g. head of
laboratory)
3 0022
4 0198 Service technicians
Tab. 1: Access level
8.2 List of parameters
No.
Parameter Description Access
level
Range of values Factory
setting
1 SterTemp
(Cycle 13)
Sterilisation temperature 1 30.0 – 140.0 °C
(opt. 150.0)
Depends
on program
2 SterTime
(Cycle 14)
Sterilisation time 1 1 – 9999 min Depends
on program
3 DryTime
(Cycle 55)
Drying time 1 1 – 9999 min Depends
on program
4 Pulses
(Cycle 42)
No. of steam/vacuum pulses in the pre-
vacuum phase:
Setting of number of steam pulses for
segmented heating
3 0 – 9999 Depends
on program
5 EndTemp
(Cycle 34)
Unloading temperature
As long as the temperature on
“CoolPT100” is higher than the set value,
the cover will stay locked.
2 Liq.:
30.0 – 80.0 °C
Drop: 99.0 °C
Solids: 120.0 °C
Depends
on program
6 SterTimeDays
(Cycle 75)
Sterilisation time in days (optional) 3 0 – 99 days 0 days
7 PulseVac1
(Cycle 45)
Vacuum value of first vacuum pulse 3 With vacuum:
0.0 – 150.0 kPa
Without vacuum:
120.0 – 150.0 kPa
Depends
on program
8 PulseVacT1
(Cycle 46)
Run-on time of first vacuum pulse 3 0 – 9999 sec. Depends
on program

Table of Contents