Revision: 2.1 8-3
No.
Parameter Description Access
level
Range of values Factory
setting
9 PulsePress1
(Cycle 47)
Steam pressure for the first steam pulse 3 With vacuum:
90.0 – 999.0 kPa
Without vacuum:
120.0 – 999.0 kPa
Depends
on program
10 PulseVac2
(Cycle 48)
Vacuum value of the second and
subsequent vacuum pulses
3 With vacuum:
0.0 – 150.0 kPa
Without vacuum:
120.0 – 150.0 kPa
Depends
on program
11 PulseVacT2
(Cycle 49)
Run-on time for the second and
subsequent vacuum pulses
3 0 – 9999 sec. Depends
on program
12 PulsePress2
(Cycle 50)
Steam pressure for the second and
subsequent steam pulses
3 With vacuum:
90.0 – 999.0 kPa
Without vacuum:
120.0 – 999.0 kPa
Depends
on program
13 PulseVac3
(Cycle 51)
Vacuum value of last vacuum pulse 3 With vacuum:
0.0 – 150.0 kPa
Without vacuum:
120.0 – 150.0 kPa
Depends
on program
14 PulseVacT3
(Cycle 52)
Run-on time of last vacuum pulse 3 0 – 9999 sec. Depends
on program
15 PulsePress3
(Cycle 53)
Steam pressure for the last steam pulse 3 With vacuum:
90.0 – 999.0 kPa
Without vacuum:
120.0 – 999.0 kPa
Depends
on program
16 ExShootOn
(Cycle 39)
This parameter allows the steam release
through the drain valve to be pulsed if
ExhMode (parameter 26) is set from 1 to
3, or is set to 4 and no compressed air is
connected
3 0 – 100 x 0.1 sec.
Depends
on program
17 ExShootOff
(Cycle 40)
3 0 – 100 x 0.1 sec.
Depends
on program
18 HoldTemp
(Cycle 57)
Hold temperature to be maintained in the
pressure chamber after the sterilisation
cycle (only for liquid programs) if
HoldTime > 0
3 40.0 – 100.0 °C 80 °C
19 HoldTime
(Cycle 76)
Time for which the product is held at the
HoldTemp after reaching the EndTemp:
0 = Function inactive
1 = Time is infinite (until the “STOP”
button is pushed with the door open)
2 = Hold function using steam valve,
max. 24 hours (X devices only)
> 2 = Hold time in minutes
3 0 – 9999 min. 0
20 AutoOpenDoor
(Cycle 68)
Defines whether the cover should open
automatically after the end of the
program (if no fault has occurred):
0 = Cover stays closed
1 = Cover opens automatically
3 0.1 0