Technical Reference
TA I
NSTRUMENTS
TGA 20504–16
Table 4.1
Method Status Codes
(continued)
Code Meaning
Cooling The heater is cooling, as
specified by a Ramp
segment.
Ending The method is complete
and the furnace is cooling
until it can Air Cool or
open and unload.
Equilb The temperature is being
equilibrated to the desired
set point.
Heating The heater temperature is
increasing, as specified by
a Ramp segment.
Holding Thermal experiment
conditions are holding;
the program is suspended.
Select Resume to con-
tinue the run.
Hot The temperature is
beyond the set point, and
the instrument cannot
remove heat fast enough
to follow the thermal
program. This is usually
caused by a large ballistic
jump to a lower tempera-
ture.
(table continued)