Installing the Hang-Down Wires
Installing the
Hang-Down Wires
During installation, take care not to bend the
hang-down wires or damage the hang-down loops.
1. Turn on the TGA instrument (see Chapter 3
for proper start-up order).
2. Press the FURNACE key to lower the
furnace.
3. Locate the sample hang-down wire in your
TGA Accessory Kit.
4. Hold the wire in your hand so that the
doubly bent top hook is pointing to the left
and the bottom hook is pointing to the right.
5. Carefully insert the bottom of the hang-
down wire into the top of the furnace far
enough so that you can insert the top of the
wire into the hang-down tube without
bending the wire (Figure 2.14 on the next
page).