Installing the TGA 2050
2–22 TA I
NSTRUMENTS
TGA 2050
4. Using tweezers, remove the foam insert
from around the screw hole (Figure 2.13):
a. Gently compress the foam with the
tweezers, being careful not to touch the
balance.
b. Remove the foam insert from the
balance chamber.
Figure 2.13
Removal of
Foam Insert
from Balance Chamber
5. Replace the sample side cover and screw.
6. Repeat the procedure to remove the foam
insert in the tare (right) side of the balance.
Foam
Insert