Maintenance & Diagnostics
5–6
Figure 5.1
Cleaning the
Inside of the
Funace Tube
6. Rinse the furnace tube with the solvent
again.
7. Replace the rubber cap on the quartz tube
stem when you have completed the cleaning
procedure.
8. Purge the system with nitrogen for one hour.
9. Heat the furnace to 900°C to remove any
remaining solvent.