Figure 2.5
Heat Exchanger Bottle
2. Pour one half bottle of TA Instruments TGA
Conditioner (PN 952377.001) into the water
reservoir bottle. Then fill the bottle to the
inner rim (see Figure 2.5) with distilled
water.
After the system has been started, recheck the
level of water in the reservoir bottle and refill to
the inner rim with distilled water, if necessary.
Do not put any water other than distilled
water in the heat exchanger reservoir.
3. Replace and tighten the water reservoir cap.
Installing the Instrument
Fill to bottle's inner rim
NOTE:
tt
tt
t
CAUTION: