EasyManuals Logo

TA Instruments TGA 2050 User Manual

TA Instruments TGA 2050
130 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #65 background imageLoading...
Page #65 background image
TA I
NSTRUMENTS
TGA 2050 35
Calibrating the TGA
The TGA 2050 requires two kinds of calibra-
tion: temperature calibration and weight cali-
bration. Both of these calibrations should be
performed when you first install the instrument
and at least once a month thereafter. Consult the
see the online help and documentation for
further information.
Temperature
Calibration
Temperature calibration involves analyzing a
material for its Curie Temperature, and compar-
ing the observed value with the standard litera-
ture value. The standard most often used is
nickel with a curie temperature of 354.4°C
(NIST Certificate for GM761). The Curie
Temperature corresponds to the extrapolated
endpoint on the "s-shaped" thermal curve
generated utilizing a magnet.
Figure 3.1
Curie Temperature
Determination
Calibrating the TGA

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TA Instruments TGA 2050 and is the answer not in the manual?

TA Instruments TGA 2050 Specifications

General IconGeneral
BrandTA Instruments
ModelTGA 2050
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals