5. Programming Procedure
5.2 Procedure for Defining a New Program
Name:
890
ABCDE
Select required plate type
(1 - 9)
Select the type of manifold to be used
(8, 12, 16 way or 384 Manifold)
Select the dispensing speed (1 - 5 or drip)
Prog XX:
- + other yes
Select required program to be defined
Enter the required name for the program
(up to eight characters)
Plate: 1
- + exit yes
Manifold Type: 8
- + exit yes
Disp Rate: 3
- + exit yes
Asp Rate: 2
- + exit yes
Crosswise: NO
- + exit yes
Selected aspirating speed
(1 - 5
[
8 or 12 manifold
]
or 1 - 3 (16 manifold or 384 Well
Manifold)
The speed of the aspirating pump.
Select if crosswise aspiration is to be used
Crosswise aspiration is only for flat bottomed plates.
Plate type: flat bottomed
Plate type:
round bottomed
Select required operating mode
(strip or plate)
Select required process step
C:1 P:1 END
- + other yes
Mode: Plate
- + exit yes
Sel. Strips: NO
- + exit yes
Select if strips are to be selected
Aspirate Aspirate solution out of the wells.
Wash Aspirate and dispense a solution
Dispense Dispense a solution
Soak Leave the solution in the wells for a time with or without shaking
End End the programming procedure for this cycle.
The speed of the dispensing pump.
5-2 Instructions for Use for Columbus Washer No. I 109 006 Rev No. 2.2 May 2003