3 - Technical Data
Emissions
Tecan Vacuum Separator (Te-VacS™) Operating Manual, 391236, en, V4.0 19 / 54
3.6.2 Chemical Resistance to Vapors
(1) resistant, (2) resistant to a limited extend, (3) not resistant
Parts
DMSO
Acetonitrile
Methanol
Methylene
Chloride
Hexane,
Chloroform
DMF, Ethyl
Acetate
Ethanol,
Isopropanol
Sodium
Hypochlorite, 10%
Waste bottle
sealing
2 1 1 3 3 2 1 1
Vacuum con-
trol valve:
Sealing,
O‑rings
1 3 1 2 1 3 1 1
Pressure sen-
sor
3 3 2 3 3 3 1 2
Table3: Parts with contact to vapors
3.6.3 Foam Formation
Substances causing foam formation should be avoided. Please note that often
protein solutions can be a source of foam.
If foam formation cannot be avoided, it is recommended to install a second waste
bottle with sufficient tubing.
Refer to “Connecting an Additional Waste Bottle” [
}
31].
3.7 Emissions
Refer to the Operating Manual of the instrument. Refer to section “Reference
Documents” [}8].
3.8 Software Requirements
The Te-VacS may be operated by the following software packages:
Freedom EVOware
FluentControl
Other Tecan software packages supporting Te-VacS
It is strongly recommended to use the latest software version. For further
information please consult the “Customer Support” [
}
51].