7 - Troubleshooting
Troubleshooting Table
Tecan Vacuum Separator (Te-VacS™) Operating Manual, 391236, en, V4.0 43 / 54
Problem/Error Possible Cause Corrective Measure
Gripping device
access failed
Wrong adapter Use the adapter that fits to the
SPE cartridge.
Gripping device
crash
Positioning pins or neighbor-
ing carriers block the arm ac-
cess
Vector setpoints for the arm
incorrect
Check the positioning pins.
Check the distance to the neigh-
boring carrier on the worktable.
Check the vector setpoints for the
arm.
Refer to the operating manual or/
and the instrument software man-
ual of the Tecan instrument used.
Decreasing pump
capacity
Salts crystallized in the pump
head
If possible, use solutions with
lower salt concentration.
Please consult the “Customer
Support” [}51].
Table7: Te-VacS troubleshooting table