- 26-
ERROR
code
Description
Apparatus
status
Reason Solution
Error 0
(Wrong EERPOM
Data)
Off Status Wrong EEPROM Data
Appoint special personnel to
adjust the apparatus
Error 1 (Too strong light)
Off Status
Optical system is not
well
Appoint special personnel to
adjust optical system of the
apparatus
Error 3
(Improper Strip
placement)
The
apparatus
does not do
testing
Misplaced strip
Rightly place the test
strip
Error 4
(Weak light
intension)
Off status
Optical system of the
apparatus is not well
Abnormal sample color
Misplaced strip
Dirty test board
Appoint special
personnel to adjust the
optical system of
apparatus
Replace the sample and
do it again
Rightly place the strip
Clean the test board
Error 7
(Incorrect strip
type)
Pause status
Use other type strip
Strip color is too deep
The apparatus is not
well.
Use the stipulated type
test paper
Verify the voltage of
power supply
Error 8
(Cannot test
continually)
Off Status
The test board is
befouled
Continually prompt 6
malfunctions but not
solve them
Clean the test board
Solve malfunctions
prompted before
Error 9
(Lose control in the
motor limitation)
The
apparatus do
not work
Inappropriate test board
position
Adjust the position of the
board
Error B
(Wrong test paper
labels
The
apparatus do
not do testing
Use other types of test
paper.
Wrong test paper labels)
Use the stipulated types
of test paper