Maintenance-314  Service 
NOTE 
The spring tension of the pin socket ensures a good 
connection between the circuit board and the pin. 
This spring tension can be destroyed by using the 
socket  as  a  connecting  point  for  spring-loaded 
probe tips, alligator clips, etc. 
Lead-End Pin Connectors. 
The pin connectors used to 
connect the leads to the interconnecting pins arecrimped 
to the ends of the leads. To remove and replace damaged 
lead-end pin connectors, remove the damaged connector 
from  the  end  of  the  lead  and  crimp  the  replacement 
connector to the lead. 
Some of the pin connectors are grouped and mounted 
in a plastic  holder. These  connectors are removed and 
installed in the instrument as a multi-pin connector unit. 
To provide correct orientation of the multi-pin connector 
with the circuit board, an arrow is stamped on the circuit 
board and a matching arrow  is  molded into the plastic 
housing of  the connector  body. Align the arrows when 
connecting to the circuit board. 
If individual lead-end pin connectors are removed from 
the plastic holder, note the wire color and position of the 
individual wires. 
Recalibration After  Repair 
After  any  electrical  component  has  been  replaced, 
recheck the calibration of the affected circuit as well as the 
calibration of other closely related circuits. Since the low- 
voltage supplies affect all circuits, calibration of theentire 
instrument should be checked if repairs have been made 
to the power supply. 
Instrument Repackaging 
If the instrument is to be shipped for long distances by 
commercial transportation, it is recommended that  it be 
repackaged in the original manner for maximum protec- 
tion. Save the original shipping carton for this purpose.