Commands Listed in Alphabetical Order
TRIGGER:CLKRE
C:CRC:CLOCKOUTPUT?
might return
:TRIGGER:CLKREC:CRC:CLOCKO UTPUT 1, indicating that the clock output
is enabled.
NOTE. 0 is returned if a Clock Recovery instrument is not present.
TRIGger:CLKRec:CRC:CREATESTandard (No Query Form)
This command saves a new clock recovery standard to the non-volatile memory
of the connected Clock Recovery instrument. Use this command to name the
new standard which is based on the current settings for NOMFREQUENCY,
LOCKRANGE, LOOPBANDWITH, PEAKING, and NOMEDGEDENSITY.
Group
Trigger
Syntax
TRIGger:CLKRec:CRC:CREATES Tandard <Qstring>
Related Commands
TRIGger:CLKRec:CRC:NOMFrequency
TRIGger:CLKRec:CRC:LOCKRange
TRIGger:CLKRec:CRC:LOOPBandwidth
TRIGger:CLKRec:CRC:PEAKing
TRIGger:CLKRec:CRC:NOMEDgedensity
TRIGger:CLKRec:CRC:DELETESTandard
Arguments
QString is a quoted string denes the le name to save the new standard. The
lename is case sensitive and spaces are allowed in the lename. Do not use
wild card characters.
Examples
TRIGGER:CLKREC:CRC:CREATES TANDARD "USER_10G" saves a new standard
named USER_10G to the non-volatile memory of the Clock Recovery instrument.
TRIGger:CLKRec:CRC:DATARate? (Query Only)
This query o nly command returns the measured data rate (in Hz) from the
connected Clock Recovery instrument.
Group
Trigger
DSA/CSA/TDS8X00/B Series Programmer Manual 2-311