Appendix H: List of Beta Commands and Features
Arguments:
<NR3> -Frequen
cy value for modulation in Hz from 100.0E-3 to 50.0E+6 with a
resolution of 100.0E-3. Maximum value is frequency maximum specified by the
shape chosen using AFG:FUNCtion.
Examples:
AFG:MODULation:FM:FREQuency 10.0E+3 sets the modulation frequency to
10kHz.
AFG:MODULation:FM:WAVeform
Description:
Sets the modulation shape to use for Frequency Modulation for the AFG.
Conditions:
Beta Command, see conditions of use. Default setting is SINE.
Group:
AFG
Syntax:
AFG:MODULation:FM:WAVeform
{SINE|SQUare|TRIangle|RAMP|NRAMP|NOISe }
AFG:MODULation:FM:WAVeform?
Arguments:
SINE -Sinewaveform
SQUare - Square waveform
TRIangle - Triangle waveform
RAMP - Positive ramp waveform
NRAMP - Negative ramp waveform
NOISe - Noise waveform
Examples:
AFG:MODULation:FM:WAVeform SINE sets the modulation waveform to SINE.
AFG:MODULation:TYPe
Description:
Sets the modulation type for the AFG. AFG:RUNMode must be set for types
to be set. The command will return NONe unless AFG:RUNMode
is set to
MODUlation.
Conditions:
Beta Command, see conditions of use. Default setting is NONe.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual H-5

Table of Contents

Other manuals for Tektronix MDO4000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO4000B Series and is the answer not in the manual?

Tektronix MDO4000B Series Specifications

General IconGeneral
BrandTektronix
ModelMDO4000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals