EasyManua.ls Logo

Tektronix MDO4000B Series User Manual

Tektronix MDO4000B Series
1158 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #151 background imageLoading...
Page #151 background image
Commands Listed in Alphabetical Order
ACTONEVent:A
CTION:SRQ:STATE
Sets or returns the enabled state of the generate service request (SRQ) action
when a specified event occurs. The default state is 0.
When this control is set, the instrument will set the OPC bit (bit 0) in the SESR
(Standard Events Status Register) when the specified event occurs. In orderto
generate a s
ervice request interrupt (SRQ), theSRER(ServiceRequestEnable
Register), DESER (Device Event Status Enable Register), and the ESER (Event
Status Enable Register) must be set appropriately as described in the programmer
manual Synchronization Methods section. Upon completion of the mask test,
event code 2600 ("Mask testing complete") is queued to the event queue.
NOTE. With this setting enabled, a *OPC? query will not return until the specied
event oc
curs. Users should be aware of this when using synchronization methods
such as *OPC? and *WAI for other operations.
To spec
ify an event, use the command ACTONEVent:EVENTTYPe.
Group
Act on Event
Syntax
ACTONEVent:ACTION:SRQ:STATE <0|1|OFF|ON>
ACTONEVent:ACTION:SRQ:STATE?
Arguments
1, ON generates an SRQ when the specified event occurs.
0, OFF turns off this action.
Examples
ACTONEVENT:ACTION:SRQ:STATE ON specifies to generate an SRQ when the
specified event occurs.
ACTONEVent:ACTION:STOPACQ:STATE
Stops acquisitions when a specified event occurs, or queries the state of the “stop
acquisition” action. The default state is 0 (off).
To specify an event, use the command ACTONEVent:EVENTTYPe.
G
roup
Act on Event
Syntax
ACTONEVent:ACTION:STOPACQ:STATE <0|1|OFF|ON>
ACTONEVent:ACTION:STOPACQ:STATE?
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-121

Table of Contents

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO4000B Series and is the answer not in the manual?

Tektronix MDO4000B Series Specifications

General IconGeneral
Bandwidth100 MHz to 1 GHz
Form FactorBenchtop
Vertical Resolution8 bits
Input CouplingAC, DC, GND
Timebase Accuracy± 1 ppm
Operating Temperature0 °C to +50 °C
Built-in Spectrum AnalyzerYes
Channelsup to 4 analog channels
Record Lengthup to 20 Mpoints
Display10.4 inch (264 mm) color TFT LCD
ConnectivityUSB, LAN, GPIB (optional)
RF Capture BandwidthUp to 1 GHz
Arbitrary/Function Generatoroptional
Input Impedance1 MΩ || 13 pF
Trigger TypesEdge, Pulse Width
Protocol AnalysisI2C, SPI, CAN, LIN, RS-232/422/485/UART, USB 2.0, Ethernet, FlexRay, MIL-STD-1553