Commands Listed in Alphabetical Order
BUS:B<x>:LIN
:IDFORmat
Specifies the LIN bus ID format.
Conditions
This command requires a DPO4AUTO or DPO4AUTOMAX application
module for all 4000 series models, and the MDO3AUTO application module
for MDO3000 series models.
Group
Bus
Syntax
BUS:B<x>:LIN:IDFORmat {NOPARity|PARity}
BUS:B<x>:LIN:IDFORmat?
Arguments
NOPARity sets the LIN bus ID format to no parity.
PARity sets the LIN bus ID format to parity.
Examples
BUS:B1:LIN:IDFORMAT PARITY sets the LIN bus ID format to parity.
BUS:B1:LIN:IDFORMAT? might return BUS:B1:LIN:IDFORMAT NOPARIT Y
indicating the LIN bus ID format is no parity.
BUS:B<x>:LIN:POLarity
Specifies the LIN bus polarity.
Conditions
This command requires a DPO4AUTO or DPO4AUTOMAX application
module for all 4000 series models, and the MDO3AUTO application module
for MDO3000 series models.
Group
Bus
Syntax
BUS:B<x>:LIN:POLarity {NORMal|INVerted}
BUS:B<x>:LIN:POLarity?
Arguments
NORMal specifies normal polarity.
INVerted specifies inverted polarity.