Commands Listed in Alphabetical Order
Examples
BUS:B1:MIL155
3B:RESPonsetime:MAXimum
12.0E-6 specifies the maximum
response time to a valid command received to be 12.0000E-6 microseconds.
BUS:B1:MIL15
53B:RESPonsetime:MAXimum?
might return 12.0000E-6.
BUS:B<x>:MIL1553B:RESPonsetime:MINimum
This command specifies the minimum response time to a valid command issued
for the MIL-STD-1553 bus.
NOTE. The MIL-STD-1553 bus specification requires devices to respond to a
valid command within 4 to 12 microseconds.
Conditions
This command requires a DPO4AERO application module for 4000 series models,
and an MDO3AERO application module for MDO3000 series models.
Group
Bus
Syntax
BUS:B<x>:MIL1553B:RESPonsetime:MINimum <NR3>
BUS:B<x>:MIL1553B:RESPonsetime:MINimum ?
Related Commands
BUS:B<x>:MIL1553B:RESPonsetime:MAXimum
Arguments
<NR3> is a floating point number that specifies the minimum response time in
seconds.
Examples
BUS:B1:MIL1553B:RESPonsetime:MINimum 4.0E-6 specifies the minimum
responsetimetoavalidcommandreceivedto4.0E-6microseconds.
BUS:B1:MIL1553B:RESPonsetime:MINimum? might return 4.0000E-6.
BUS:B<x>:MIL1553B:SOUrce
This command specifies the source for differential input for the MIL-STD-1553
bus. The supported source waveforms are channels 1–4, math waveform, and
reference waveforms 1–4. The default is channel 1.
Conditions
This command requires a DPO4AERO application module for 4000 series models,
and an MDO3AERO application module for MDO3000 series models.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-187