Commands Listed in Alphabetical Order
BUS:B<x>:SPI
:IDLETime
This command specifies the idle time, in seconds, for the SPI bus.
Conditions
This command requires a DPO4EMBD application module for 4000 series
models, and an MDO3EMBD application module for MDO3000 series models.
Group
Bus
Syntax
BUS:B<x>:SPI:IDLETime <NR3>
BUS:B<x>:SPI:IDLETime?
Arguments
NR3 is the idle time, in seconds, for the SPI bus.
Examples
BUS:B1:SPI:IDLETIME 100.0000E-9 sets the idle time to 100 ns.
BUS:B1:SPI:IDLETIME? might return BUS:B1:SPI:IDLE TIME
100.0000E-9
indicating the idle time is set to 100 ns.
BUS:B<x>:SPI{:SELect|:SS}:POLarity
This command specifies the polarity for the SPI bus.
Conditions
This command requires a DPO4EMBD application module for 4000 series
models, and an MDO3EMBD application module for MDO3000 series models.
Group
Bus
Syntax
BUS:B<x>:SPI{:SELect|:SS}:POLarity {LOW|HIGH}
BUS:B<x>:SPI{:SELect|:SS}:POLarity?
Arguments
LOW specifies an active low polarity.
HIGH specifies an active high polarity.
BUS:B<x>:SPI{:SELect|:SS}:SOUrce
This command specifies the source waveform for the SPI bus.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-199