EasyManua.ls Logo

Tektronix MDO4000B Series

Tektronix MDO4000B Series
1158 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Commands Listed in Alphabetical Order
(See page 2-61, Se
arch Command Group.)
Returns
<NR1> = 1 if the CAN bus feature is present.
<NR1> = 0 if the CAN bus feature is not present.
CONFIGuration:BUSWAVEFORMS:CANFD? (Query Only)
This query returns a Boolean value to indicate whether the optional CAN FD bus
triggering and analysis feature is present.
Conditions
This feature requires installation of a DPO4AUTO or DPO4AUTOMAX
application module for MDO4000C models, or MDO3AUTO for MDO3000
models.
Group
Configuration
Syntax
CONFIGuration:BUSWAVEFORMS:CANFD?
Related Commands
(See page 2-15, Bus Command Group.)
(See page 2-76, Trigger Command Group.)
(See page 2-61, Search Command Group.)
Returns
<NR1>: = 1 if the CAN FD bus feature is present.
<NR1>: = 0 if the CAN FD bus fe
ature is not present.
CONFIGuration:BUSWAVEFORMS:ETHERNET? (Query Only)
This query returns a boolean value to indicate whether the optional Ethernet
triggering and analysis feature is present. (Enabling this feature on MDO4000/B/C
and MSO/DPO4000B requires installation of a DPO4ENET application module.)
Group
Configuration
Syntax
CONFIGuration:BUSWAVEFORMS:ETHERNET?
Related Commands
(See page 2-15, Bus Command Group.)
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-245

Table of Contents

Related product manuals