Commands Listed in Alphabetical Order
<NR1> is an inte
ger. 0 turns off this feature; any other integer turns it on.
Examples
MASK:TEST:SRQ:FAILURE? might return :MASK:TEST:SRQ:FAILURE 1,
indicating that the pass/fail SRQ on failure is turned on.
MASK:TESt:STATE
This command turns the pass/fail mask test on or off. Most of the other
MASK:TEST commands need to be executed before this command.
NOTE. To ensure that Mask commands are enabled, use the command
APPLication:TYPe LIMITMask.
A series of examples showing how to use mask commands for typical tasks
is included in an appendix. (See page E-1, Mask/Limit Command Sequence
Examples.)
Conditions
This command requires a DPO4LMT application module for 4000 series models,
and an MDO3LMT application module for MDO3000 series models.
Group
Mask
Syntax
MASK:TESt:STATE {ON|OFF|<NR1>}
MASK:TESt:STATE?
Related Commands
MASK:TESt:WAVEform
Arguments
ON turns the mask test on.
OFF turns the mask test off.
<NR1> is an integer. 0 turns the mask test off; any other integer turns it on
.
Examples
MASK:TESt:STATE ON starts the mask test.
MASK:TESt:STATus? (Query Only)
This command indicates the result of a pass/fail mask test.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-393

Table of Contents

Other manuals for Tektronix MDO4000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO4000B Series and is the answer not in the manual?

Tektronix MDO4000B Series Specifications

General IconGeneral
BrandTektronix
ModelMDO4000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals