Commands Listed in Alphabetical Order
Examples
MASK:TESt:STO
P:FAILure 50000
sets the instrument to change the test status
to “failing” after 50,000 failed waveforms.
MASK:TESt:ST
OP:FAILure?
might return INFINITE which indicates the
instrument has been set so that there is no limit to tested waveform violations
per test.
MASK:TESt:
TIME
This command specifies the duration, in seconds, the instrument should run a
pass/fail mask test. The default is INFINITE.
A series of examples showing how to use mask commands for typical tasks
is included in an appendix. (See page E-1, Mask/Limit Command Sequence
Example
s.)
Conditions
This command requires a DPO4LMT application module for 4000 series models,
and an MDO3LMT application module for MDO3000 series models.
Group
Mask
Syntax
MASK:TESt:TIME {<NR1>|INFInite}
MASK:TESt:TIME?
Related Commands
MASK:TESt:STATE,
MASK:TESt:DELay
Arguments
<NR1> is an integer that represents the number of seconds to test. The maximum
number of seconds that can be specified is 172,800.
INFInite indicates that there is no time limit on the test run.
Examples
MASK:TESt:TIME 2 specifies to run a mask test for 2 seconds.
MASK:TESt:TIME? might return INFINITE, indicating there is currently no
specified time limit. This is the default.
MASK:TESt:WAVEform
This command specifies the number of waveforms the instrument should test
during a pass/fail mask test.
2-396 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual