Commands Listed in Alphabetical Order
Group
Measurement
Syntax
MEASUrement:MEAS<x>:SOUrce<x>
{CH1|CH2|CH3|CH4|MATH|R1|R2|R3|R4|D0|D 1|D2|D3|D4|D5|D6|D7
|D8|D9|D10|D11|D12|D13|D14|D15|HIStogr am|RF_AMPlitude|
RF_FREQuency|RF_PHASe}
MEASUrement:MEAS<x>:SOUrce<x>?
Related Commands
MEASUrement:MEAS<x>:TYPe
Arguments
CH1–4 is an analog channel to use as the source waveform.
MATH is the math waveform.
REF1–4 is a reference waveform to use as the source waveform.
D0–D15 is a digital channel to use as the source waveform.
(MSO/MDO4000/B models only as well as MDO3000 and MDO400C models
with option MDO3MSO or MDO4MSO installed.)
HIStogram indicates the histogram is the object to be measured. HIStogram only
applies to SOUrce1; it is not allowed on SOUrce2.
RF_AMPlitude|RF_FREQuency|RF_PHASe are the RF time domain traces.
(MDO4000/B/C models only.)
Examples
MEASUREMENT:MEAS4:SOURCE2 CH1 specifies CH1 as the delay "to" source
when making delay measurement.
MEASUREMENT:MEAS2:SOURCE2? might return
:MEASUREMENT:MEAS2:SOURCE2 MATH1 indicating that Math 1 is the
measurement 2 source.
MEASUrement:MEAS<x>:STATE
This command specifies whether the specified measurement slot is computed
and displayed. The measurement slot is specified by <x>, which ranges from
1 through 8.
For a measurement to display, you must have selected a source waveform
and defined the measurement you want to take and display. You select the
measurement using the MEASUrement:MEAS<x>:SOUrce<x> command.
You define the measurement type using the MEASUrement:MEAS<x>:TYPe
command.
Group
Measurement
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-437