Commands Listed in Alphabetical Order
Arguments
<NR3> isafloati
ng point number that specifies the IEC observation period. Valid
values ranges from 0.2 to 10 s.
Examples
POWER:HARMONICS:IEC:OBSPERIOD 3 sets the IEC observation period to
3.0000 s.
POWER:HARMONICS:IEC:OBSPERIOD? might return
:POWER:HARMONICS:IEC:OBSPERIOD 10 indicating that the IEC observation
period is set to 10 s.
POWer:HARMonics:IEC:POWERFACtor
This command specifies the rated power factor for IEC harmonics.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:HARMonics:IEC:POWERFACtor <NR3>
POWer:HARMonics:IEC:POWERFACtor?
Arguments
<NR3> is a floating point number that specifies the power factor. Valid values
ranges from 0 to 1 in increments of 0.1.
Examples
POWER:HARMONICS:IEC:POWERFACTOR 1 sets the power factor to 1.0000.
POWER:HARMONICS:IEC:POWERFACTOR? might return
:POWER:HARMONICS:IEC:POWERFACTOR 900.0000E-03 indicating
that the power factor is set to 0.9.
POWer:HARMonics:MIL:FUNDamental:CALCmethod
This command specifies the measurement method for the MIL harmonics
fundamental current for use in calculating limits.
Conditions
This command requires a DPO4PWR application module for 4000 serie
smodels,
or an MDO3PWR application module for MDO3000 series models.
Group
Power