Commands Listed in Alphabetical Order
Syntax
POWer:HARMoni
cs:MIL:FUNDamental:CALCmethod {MEAS|USER}
POWer:HARMonics:MIL:FUNDamental:CALCme thod?
Arguments
MEAS specifies that the value of the fundamental current used in calculating limits
is measured.
USER specifies that the value of the fundamental current used in calculated limits
is user defined.
Examples
POWER:HARMONICS:MIL:FUNDAMENTAL:CALCME THOD USER sets the
measurement method to User.
POWER:HARMONICS:MIL:FUNDAMENTAL:CALCME THOD? might return
:POWER:HARMONICS:MIL:FUNDAMENTAL:CALCM ETHOD MEAS indicating that
the measu
rement method is set to Meas.
POWer:HARMonics:MIL:FUNDamental:USER:CURrent
This command specifies RMS amperes for USER CALCmethod.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:HARMonics:MIL:FUNDamental:USER:C URrent <NR3>
POWer:HARMonics:MIL:FUNDamental:USER:C URrent?
Arguments
<NR3> is a floating point number that specifies the current in amperes for USER
CALCmethod.
Examples
POWER:HARMONICS:MIL:FUNDAMENTAL:USER:C URRENT 2 sets the RMS
fundamental current to 2 A in calculating limits.
POWER:HARMONICS:MIL:FUNDAMENTAL:USER:C URRENT? might return
:POWER:HARMONICS:MIL:FUNDAMENTAL:USER: CURRENT 1.0000 indicating
that the RMS fundamental current for use in calculating limits is 1.0000 A.
POWer:HARMonics:MIL:LINEFREQuency
This command specifies the line frequency for MIL-STD-1399 Section 300A
harmonics tests. Valid values are 60 or 400 Hz.
2-474 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual