Commands Listed in Alphabetical Order
POWer:HARMon
ics:RESults:IEC:FUNDamental? (Query Only)
Returns the IEC fundamental current used in calculating limits.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:HARMonics:RESults:IEC:FUNDamenta l?
Examples
POWER:HARMONICS:RESULTS:IEC:FUNDAMENTA L? might return
:POWER:HARMONICS:RESULTS:IEC:FUNDAMENT AL 1.0 indicating that the
IEC fundamental frequency is set to 1.0 A.
POWer:HARMonics:RESults:IEC:HARM3ALTernate? (Query Only)
Returns the IEC harmonics test result for the 3rd harmonic.
NOTE. The query returns NA if the limit does not apply.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:HARMonics:RESults:IEC:HARM3ALTer nate?
Returns
PASS, FA IL,orNA.
Examples
POWER:HARMONICS:RESULTS:IEC:HARM3ALTER NATE? might return
POWER:HARMONICS:RESULTS:IEC:HARM3ALTER NATE PASS indicating the
test result.
POWer:HARMonics:RESults:IEC:HARM5ALTernate? (Query Only)
Returns the overall harmonics test result for the 5th harmonic.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-481