Commands Listed in Alphabetical Order
NOTE. The query
returns NA the harmonic limit does not apply.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR
application module for MDO3000 series models.
Group
Power
Syntax
POWer:HARMonics:RESults:IEC:HARM5ALTer nate?
Returns
PASS, F AIL,orNA.
Examples
POWER:HARMONICS:RESULTS:IEC:HARM5ALTER NATE? might return
POWER:HARMONICS:RESULTS:IEC:HARM5ALTER NATE PASS indicating the
test result.
POWer:HARMonics:RESults:IEC:POHC? (Query Only)
Returns the IEC POHC measurement.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:HARMonics:RESults:IEC:POHC?
Examples
POWER:HARMONICS:RESULTS:IEC:POHC? might return
:POWER:HARMONICS:RESULTS:IEC:POHC 0.5 indicating that the
IECPOHCissetto0.5A.
POWer:HARMonics:RESults:IEC:POHL? (Query Only)
Returns the IEC POHL measurement.
Conditions
This command requires a DPO4PWR application modu
le for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
2-482 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual