Commands Listed in Alphabetical Order
Conditions
This command re
quires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:REFLevel:PERCent:HIGH <NR3>; Ranges={D,0.0,100.0}
POWer:REFLevel:PERCent:HIGH?
Arguments
<NR3> is a floating point number that specifies the high value in percent.
Examples
POWER:REFLEVEL:PERCENT:HIGH 95 sets the high reference level to 95% of
the signal amplitude.
POWER:REFLEVEL:PERCENT:HIGH? might return
:POWER:REFLEVEL:PERCENT:HIGH 90 indicating that the percentage high
reference level is set to 90% of the signal amplitude.
POWer:REFLevel:PERCent:LOW
This command specifies the low reference percent level to be used for power
measurements.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:REFLevel:PERCent:LOW <NR3>; Ranges={D,0.0,100.0}
POWer:REFLevel:PERCent:LOW?
Arguments
<NR3> is a floating point number that specifies the low value in percentage.
Examples
POWER:REFLEVEL:PERCENT:LOW 15 sets the low reference level to 15% of the
signal amplitude.
POWER:REFLEVEL:PERCENT:LOW? might return
:POWER:REFLEVEL:PERCENT:LOW 10 indicating that the percentage low
reference level is set to 90% of the signal amplitude.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-503