Commands Listed in Alphabetical Order
POWer:REFLev
el:PERCent:MID<x>
This command specifies one of 3 mid reference percentage levels to be used for
power measurements. MID1 is used on the users voltage waveform. MID2 is
used on the us
ers current waveform. MID3 is used on the users gate waveform.
(MID3 is specific to power applications.)
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:REFLevel:PERCent:MID<x> <NR3>; Ranges={D,0.0,100.0}
POWer:REFLevel:PERCent:MID<x>?
Arguments
<NR3> is a floating point number that specifies the mid value in percentage.
Examples
POWER:REFLEVEL:PERCENT:MID1 25 sets the mid reference voltage to 25%.
POWER:REFLEVEL:PERCENT:MID1? might return
POWER:REFLEVEL:PERCENT:MID1 50.0000 indicating that the mid reference
voltage level is set to 50%.
POWer:RIPPle (No Query Form)
This command performs a vertical autoset for ripple measurements or sets the
vertical offset to 0.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:RIPPle {VERTAUTOset|VERTDEFault}
Arguments
VERTAUTOset automatically scales the source waveform to optimize ripple
measurements.
VERTDEFault sets the vertical offset of the source waveform to 0 volts (for
voltage source) or 0 amperes (for cur
rent source).
2-504 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual