Commands Listed in Alphabetical Order
Examples
POWER:SOA:LOG
:YMAX 10
sets the YMAX value to 10.0000.
POWER:SOA:LOG:YMAX ? might return :POWER:SOA:LOG:YMAX 100.0000
indicating th
at the YMAX value for log plot type is set to 100.0000.
POWer:SOA:LOG:YMIN
This command specifies the user YMIN value for use in Log SOA calculations.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR applicat
ion module for MDO3000 series models.
Group
Power
Syntax
POWer:SOA:LOG:YMIN <NR3>
POWer:SOA:LOG:YMIN?
Arguments
<NR3> is a floating point number that specifies the YMIN value used for log
SOA calculations.
Examples
POWER:SOA:LOG:YMIN 1 sets the YMIN value to 1.0000.
POWER:SOA:LOG:YMIN ? might return :POWER:SOA:LOG:YMIN 10
0.0000E-3
indicating that the YMIN value for log plot type is set to 100.0000E-3.
POWer:SOA:MASK:DEFine
This command specifies the X (volts) and Y (Amps) coordina
tes of the current
SOA mask. You can specify the number of points from 2 to 10, minimum being 2.
Successive X values must be the preceding X values. The number of XY points
sent determines the value of NR_PT.
Conditions
Thi
s command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:SOA:MASK:DEFine <NR3>
POWer:SOA:MASK:DEFine?
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-511