Commands Listed in Alphabetical Order
Arguments
<NR3> is a floati
ng point number that represents the SOA mask coordinates.
Examples
POWER:SOA:MASK:DEFINE 1 0 specifies the SOA mask coordinates as
0.0E+0,0.0E+0,0.0E+0,0.0E+0,0.0E+0,30.0000,25.0000,30.0000,300.0000,2.5000,
300.0000,2.4000,300.0000,2.2000,300.0000,1.9000,300.0000,1.3000,
300.0000,0.0E+0.
POWER:SOA:MASK:DEFINE ? might return :POWER:SOA:MASK:DEFIN E
0.0E+0,30.0000,25.0000,30.0000,300.000 0,2.5000,300.0000,0.0E+0
indicating that the SOA mask coordinates are set to
0.0E+0,30.0000,25.0000,30.0000,300.0000,2.5000,300.0000,0.0E+0.
POWer:SOA:MASK:MAXAmps
This command specifies the maximum current applied to SOA mask testing.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:SOA:MASK:MAXAmps <NR3>
POWer:SOA:MASK:MAXAmps?
Arguments
<NR3> is a floating point number that specifies the maximum current applied
to SOA mask testing.
Examples
POWER:SOA:MASK:MAXAMPS 40 sets the SOA mask testing current to 40 A.
POWER:SOA:MASK:MAXAMPS ? might return :POWER:SOA:MASK:MAXAMPS
30.0000
indicating that the SOA mask testing maximum current is set to 30 A.
POWer:SOA:MASK:MAXVolts
This command specifies the maximum voltage applied to SOA mask testing.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
2-512 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual