Commands Listed in Alphabetical Order
Syntax
POWer:SOA:MAS
K:NR_Pt?
Examples
POWER:SOA:MASK:NR_PT? might return :POWER:SO A:MASK:NR_PT 4
indicating that the number of mask points defined is 4.
POWer:SOA:MASK:STATE
This command specifies the state of the mask for SOA calculations.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:SOA:MASK:STATE {OFF|LIMITS|POINTS}
POWer:SOA:MASK:STATE?
Arguments
OFF disables mask testing.
LIMITS enables mask testing based on limits specified using
POWer:SOA:MASK:MAXAmps, POWer:SOA:MASK:MAXVolts,and
POWer:SOA:MASK:MAXWatts commands
POINTS enables mask testing based on masks points defined.
Examples
POWER:SOA:MASK:STATEPOINTS enables mask testing power based on points
defined.
POWER:SOA:MASK:STATE? might return :POWER:SOA:MASK:STATE LIMITS
indicating that the mask testing is enabled based on limits.
POWer:SOA:MASK:STOPOnviol
This command specifies the enabled state of the mask stop on violation condition.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power