Commands Listed in Alphabetical Order
Examples
POWER:SWLOSS:
GATE:TURNON 3.5
sets the gate turn on level to 3.5000.
POWER:SWLOSS:GATE:TURNON ? might return :POWER:SWLOSS:GATE:TURNON
1.5
indicatin
gthatthegatingturnonlevelissetto1.5.
POWer:SWLoss:NUMCYCles? (Query Only)
Returns the number of cycles counted for the switching loss calculation.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:SWLoss:NUMCYCles? <NR3>
Examples
POWER:SWLOSS:NUMCYCLES? might return the number of cycles for
switching loss calculation, if the POWer:TYPe is set to Switching Loss
(
POWer:TYPe:SWITCHingLoss). If the query times out with the Switching
Loss power enabled, it indicates the possibility of incorrect settings and might
return the following:
*ESR?
20
ALLEV?
2202,"Measurement error, No period found;
:power:swloss:numcycles?",410,"Query INTERRUPTED"
POWer:SWLoss:RDSon
This command specifies the user RDSON value for use in switching loss
calculations when the conduction calculation method is RDSON.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:SWLoss:RDSon <NR3>
POWer:SWLoss:RDSon?
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-523