Commands Listed in Alphabetical Order
Conditions
This command re
quires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:SWLoss:REFLevel:PERCent:GATEMid <NR3>
POWer:SWLoss:REFLevel:PERCent:GATEMid?
Arguments
<NR3> is a floating point number that specifies the mid voltage reference level
in volts.
Examples
POWER:SWLOSS:REFLEVEL:PERCENT:GATEMID 30 sets the gate mid reference
voltage to 30%.
POWER:SWLOSS:REFLEVEL:PERCENT:GATEMID ? might return
:POWER:SWLOSS:REFLEVEL:PERCENT:GATEMID 50.0000 indicating that the
gate mid reference voltage is set to 50%.
POWer:SWLoss:REFLevel:PERCent:LOWCurrent
This command specifies the low current reference level used in switching loss
power measurements in percent.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:SWLoss:REFLevel:PERCent:LOWCurre nt <NR3>
POWer:SWLoss:REFLevel:PERCent:LOWCurre nt?
Arguments
<NR3> is a floating point number that specifies the low voltage reference level
perce
nt.
Examples
POWER:SWLOSS:REFLEVEL:PERCENT:LOWCURRE NT 15 sets the reference low
current to 15%.
POWER:SWLOSS:REFLEVEL:PERCENT:LOWCURRE NT ? might return
:POWER:SWLOSS:REFLEVEL:PERCENT:LOWCURR ENT 5.0000 indicating that
the reference low current is set to 5%.
2-526 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual