Commands Listed in Alphabetical Order
POWer:SWLoss
:REFLevel:PERCent:LOWVoltage
This command specifies the low voltage reference level used in switching loss
power measurements in percent.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:SWLoss:REFLevel:PERCent:LOWVolta ge <NR3>
POWer:SWLoss:REFLevel:PERCent:LOWVolta ge?
Arguments
<NR3> is a floating point number that specifies the low voltage reference level in
percent.
Examples
POWER:SWLOSS:REFLEVEL:PERCENT:LOWVOLTA GE 10 sets the reference low
voltage to 10%.
POWER:SWLOSS:REFLEVEL:PERCENT:LOWVOLTA GE ? might return
:POWER:SWLOSS:REFLEVEL:PERCENT:LOWVOLT AGE 5.0000 indicating that
the reference low voltage is set to 5%.
POWer:SWLoss:TOFF:ENERGY:MAX? (Query Only)
Returns the maximum Toff energy for the switching loss calculation.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:SWLoss:TOFF:ENERGY:MAX?
Examples
POWER:SWLOSS:TOFF:ENERGY:MAX? might return
:POWER:SWLOSS:TON:ENERGY:MAX 1 indicating that the maximum Toff energy
switching loss calculation is set to 1 J.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-527