Commands Listed in Alphabetical Order
*PSC
This command specifies the power-on status flag that controls the automatic
power-on handling of the DESER, SRER, and ESER registers. When *PSC is
true, the DES
ER register is set to 255 and the SRER and ESER registers are set
to 0 at power-on. When *PSC is false, the current values in the DESER, SRER,
and ESER registers are preserved in nonvolatile memory when power is shut off
and are restored at power-on.
Group
Status and Error
Syntax
*PSC {O FF|ON|NR1>}
*PSC?
Related Commands
DESE, *ESE, FACtory, *RST, *SRE
Arguments
OFF sets the power-on status clear flag to false.
ON sets the power-on status clear flag to true.
<NR1> = 0 sets the power-on status clear flag to false. This disables the power-on
clear allowing the oscilloscope to possibly assert SRQ after power-on; anyother
value sets the power-on status clear flag to true, enabling the power-on status clear
preventing any SRQ assertion after power on.
Examples
*PSC 0 sets the power-on status clear flag to false.
*PSC? might return 1 to indicate that the power-on status clear flag is set to true.
*PUD
This command specifies a string of Protected User Data. This data is protected
by the PASSWord command. You can modify it only by first entering the correct
password. This password is not necessary to query the data.
Group
Status and Error
Syntax
*PUD {< Block>|<QString>}
*PUD?
Related Commands
PASSWord
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-537

Table of Contents

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO4000B Series and is the answer not in the manual?

Tektronix MDO4000B Series Specifications

General IconGeneral
Bandwidth100 MHz to 1 GHz
Form FactorBenchtop
Vertical Resolution8 bits
Input CouplingAC, DC, GND
Timebase Accuracy± 1 ppm
Operating Temperature0 °C to +50 °C
Built-in Spectrum AnalyzerYes
Channelsup to 4 analog channels
Record Lengthup to 20 Mpoints
Display10.4 inch (264 mm) color TFT LCD
ConnectivityUSB, LAN, GPIB (optional)
RF Capture BandwidthUp to 1 GHz
Arbitrary/Function Generatoroptional
Input Impedance1 MΩ || 13 pF
Trigger TypesEdge, Pulse Width
Protocol AnalysisI2C, SPI, CAN, LIN, RS-232/422/485/UART, USB 2.0, Ethernet, FlexRay, MIL-STD-1553