Commands Listed in Alphabetical Order
Arguments
<Block> is a blo
ck containing up to 300 ASCII characters.
<QString> is a string containing up to 300 ASCII characters.
Examples
*PUD #229T his oscilloscope belongs to me stores the string "This
oscilloscope belongs to me" in the user protected data area.
*PUD? might return #221PROPERTY OF COMPANY X
*RCL (No Query Form)
This command restores the state of the oscilloscope from a copy of the settings
stored in memory (The settings are stored using the *SAV command).
Group
Save and Recall
Syntax
*RCL <N R1>
Related Commands
FACtory, *LRN?, RECAll:SETUp, *RST, *SAV, SAVe:SETUp
Arguments
<NR1> is a value in the range from 1 to 10, which specifies a saved setup storage
location.
Examples
*RCL 3 restores the oscilloscope from a copy of the settings stored in memory
location 3.
REBOOT (No Q uery Form)
Reboots the system after a short delay, emulating the front panel power button
push. The system will power back on after a 5 second delay.
Group
Miscellaneous
Syntax
REBOOT
RECAll:MASK (No Query Form)
This command recalls the mask from a specified file that was saved to the current
working directory using the SAVe:MASK command.
2-538 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual